Oximeter OxyGuard POLARIS with galvanic probe aquaculture france

Oximeter OxyGuard POLARIS with galvanic probe

Polaris OxyGuard Dissolved Oxygen Meter (Galvanic Sensor)

General Description

The Polaris OxyGuard is a portable dissolved oxygen meter designed for measuring oxygen concentration (O₂) in water.
It is widely used in aquaculture, laboratory work, environmental monitoring, and water quality control.

Operating Principle

The device uses a galvanic-type electrochemical sensor that produces an electric current proportional to the dissolved oxygen concentration.

  • The sensor contains two electrodes (anode and cathode) immersed in an electrolyte.

  • Oxygen diffuses through a permeable membrane and reacts at the electrode surface, generating an electrical signal.

  • The signal is then converted and displayed in mg/L or % saturation.

€0.00
Tax included
Modèle
Quantity

Oxygen Meter OXYGUARD POLARIS

Complete set includes:
Probe, battery, storage bag, accessories, and 3-meter cable (other cable lengths available upon request).

Measurement range:
Oxygen: 0.0 to 60.0 mg / l (ppm)
Oxygen saturation: 0.0 - 600%
Temperature: -5.0 - +45°C
Languages: English, German, French, Spanish
Weight with accessories: approx. 1.0 kg

The POLARIS is a portable oxygen meter. It is the best-selling model in commercial fish farming. This device is equipped with a new generation probe and microprocessor electronics. Due to its automatic calibration and automatic self-check, it is very easy to use. The POLARIS measures oxygen saturation in mg / l and % and temperature. All values are displayed simultaneously on the large backlit screen. The devices are extremely robust, waterproof, and floatable.

The POLARIS 2 with recorder, and optionally the POLARIS Link (P0012400) for data transfer to a PC.

Models

1 2
Name

OxyGuard Polaris (NEW)

P0011050

OxyGuard Polaris (OLD)

P0011310

O2 – mg/l

0.0 ... 60.0 0.0 ... 60.0
 

O2 – %

0...600 0...600
°C -5...+45 -5...+45
Précision

O2 – mg/l

+/- 1% +/- 1%

O2 – %

+/- 1% +/- 1%

°C

+/- 2°C +/- 2°C
Câble 3.0 m 3.0 m
Dimension 300-240-60 300-240-60
Weight (kg) 0.7 0.7

Technical Specifications (typical values)

  • Measurement range: 0–20 mg/L O₂ or 0–200% saturation

  • Accuracy: ±0.1 mg/L or ±1% of reading

  • Resolution: 0.01 mg/L

  • Response time (T90): approx. 30 seconds

  • Operating temperature: 0–50 °C

  • Automatic temperature compensation (built-in sensor)

  • Calibration: simple air-saturated water calibration

  • Display: digital, backlit

  • Power supply: AA batteries or rechargeable battery

  • Protection rating: IP65–IP67 depending on model

  • Probe cable: typically 3–10 m, interchangeable

Advantages

- No polarization required (unlike polarographic probes)
- Ready for use immediately after switching on
- Low power consumption
- Rugged and waterproof design
- Easy maintenance (membrane/electrolyte replacement)
- Ideal for fieldwork and fish farming ponds

Maintenance and Calibration

  • Regular cleaning of the membrane to avoid organic fouling.

  • Monthly inspection of the electrolyte and membrane integrity.

  • Air-saturated calibration before each measurement series to ensure accuracy.

Typical Applications

  • Monitoring oxygen levels in aquaculture ponds and tanks

  • Water quality assessments in lakes and rivers

  • Aeration system efficiency and biological respiration studies